| Syringe Size |
1 ml - 20 ml single-use syringe |
| Infusion Rate |
1 mm/hr - 99 mm/hr |
| Alarms |
Battery depleted, mechanical error, occlusion, syringe disengaged, completion |
| Alerts |
Battery low, syringe not loaded, near end |
| PCA Bolus |
Maximum setting of bolus is 1/4 of the syringe size |
| OLED Display |
Syringe size; rate is set by mm/hr and displayed in ml./hr simultaneously; alarm messages |
| “LOW BATT” LED indicator |
On when battery is low or depleted |
| Dimension |
166 (L) x 60 (B) x 30 (H) mm |
| Weight |
≤ 180g (battery weight not included) |
| Maximum infusion pressure |
≤ 0.12MPa(with 10 ml syringe) |
| Occlusion alarm pressure |
0.12MPa(with 10 ml syringe) |